5 Maintenance

Treads of fastening screws or the joins on the frame parts may get degreased
by disinfectants. In these cases timely lubrication prevents costly repairs.
Within the scope of executing the annual-regular revision of medical
equipment it is necessary to check the accumulator voltage (loadless). If
measrued value is lower then 12V (24V in case of measuuring of two
accumulator cell together), is necessary the replace the battery.

5.1 List of error messages and other generated messages

Description of
problem
The bed is not
responding to
controller
commands
The remote
wireless foot
controller is not
working. The
battery icon is
flashing orange.
The remote
wireless foot
controller is not
working. The
battery icon is
flashing orange.
BORCAD Medical a.s.
739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087
Likely cause
The bed probably
does not have
electricity and the
battery is exhausted.
The battery in the
wireless foot
controller is
exhausted.
The transmitter is not
paired with the
receiver.
Actions to be taken
by the operator
Plug the power supply
cord to the electric
socket (full recharging
of the battery takes 8
hours).
Replace the battery in
the remote wireless foot
controller.
Pair the transmitter with
the receiver.
Actions to be taken
by the qualified
service staff
Check, if the electric
network is live and if
the socket works.
Use some other
socket
Check, if the bed
works with a different
remote wireless foot
controller. If so,
replace it.
Check, if the bed
works with a different
remote foot
controller. If so,
replace it.
55