Maintenance - BORCAD Medical AVE2 Instruction Manual

Birthing bed
Hide thumbs Also See for AVE2:
Table of Contents

Advertisement

10 Maintenance

From practical testing results, the technical lifetime of the product is 15 years.
Lifetime of the battery is limited and must be replaced.
- batteries (these are provided with the warranty of 6 months).
- gas springs need to be replaced every 12 years.
- the upholstery is guaranteed for 2 years.
Treads of fastening screws or the joins on the frame parts may get degreased by
disinfectants. In these cases, timely lubrication prevents costly repairs.
Within the scope of annual regular revisions of a medical equipment, it is necessary
to check the battery voltage (loadless). If the measured value is below 12V (24V in
case of measuring two battery cells together), it is necessary to replace the battery
immediately.
BORCAD Medical a.s.
739 45 Fryčovice 673, tel.: +420 558 640 631
Warning:
Caution
Warning:
Dangerous electric voltage
82