2.3 Setting the product

 No objects (furniture, stands, etc.) may be placed near the bed that
could collide with it (minimum distance of 20 cm). Movable parts of the
chair must be treated with caution. Movement of these parts can cause
collision with other objects, which may lead to damage to parts of the
chair or injury of the patient, staff or any third person!
 The power cable must not be connected through an extension cord and it
must not be connected to common multisocket inlets.
 No input cable of this or other devices can lead over the surfaces of the bed
A and /or over the patient. When moving the bed avoid pinching/kinking of
any objects.
 Pay attention to proper fixing of the approved accessories listed in the
manual
 Installation is done by trained person; operating the birthing bed only
allowed by trained personnel for routine functions; operators must be trained
on setting sufficient height of the infusion stand.
 Training will be performed by a representative of BORCAD Medical a.s.
(manufacturer) or its official dealer
 Maximum weight of the patient 210 kg.
 A battery is a component of every obstetric bed; it is fully charged and
flawless and allows for the operation of the bed for the time of approx. 3
minutes. The battery is recharged automatically from the control unit and
does not require any special maintenance. If the chair is not connected to
the network for more than one week, it is necessary to disconnect the
battery connector from the CB (Control box). With internal rechargeable
battery system for all functions of the bed.