BORCAD Medical AVE2 Instruction Manual

BORCAD Medical AVE2 Instruction Manual

Birthing bed
Hide thumbs Also See for AVE2:
Table of Contents

Advertisement

BIRTHING BED AVE2
UN-PACKING MANUAL
INSTRUCTION MANUAL
D9U004PPB-0101
Version: 06
Punlication Date: 2020-11
BORCAD Medical a.s., CZ - 739 45 Fryčovice 673
T: +420 558 640 631, W: www.borcadmedical.com

Advertisement

Table of Contents
loading

Summary of Contents for BORCAD Medical AVE2

  • Page 1 BIRTHING BED AVE2 UN-PACKING MANUAL INSTRUCTION MANUAL D9U004PPB-0101 Version: 06 Punlication Date: 2020-11 BORCAD Medical a.s., CZ - 739 45 Fryčovice 673 T: +420 558 640 631, W: www.borcadmedical.com...
  • Page 2: Table Of Contents

    7.8 PPB-068 Grab handles, detachable (R/L) ............55 7.9 PPB-013 Grab handles, non-detachable, foldable (R/L) ......... 55 7.10 Holder of infusion stand ................57 7.11 Equipotential coupling ................... 57 BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 3 10.5 Transport instruction, Storage instruction ............87 10.6 Service and repairs ..................88 11 ENVIRONMENTAL PROTECTION AND LIQUIDATION OF THE PRODUCT ....... 89 12 LIST OF AVAILABLE EQUIPMENT ......................90 BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 4 Please keep these instructions near the bed throughout its operation. Every person operating the birthing bed AVE2 must read and understand the contents of these instructions. Very important information is marked throughout the manual by the following...
  • Page 5 Thermal protection of the transformer Equipotentiality Packaging material: “Keep dry” Packaging material: “Fragile!“ Packaging material: “This side up” Temperature limitation Transportation marking: „Relative air humidity“. Transportation marking: “Air pressure”. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 6 The space is indicated with a caution sticker. Warning: Do not place objects in this area Warning: Do not place objects on this cover Designation of birthing bed for adult BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 7 The power cable shall always be clamped on the selected place of the bed using a plastic hook as is recommended on picture! Follow the instructions manual! BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 8 Intermittent acoustic signal even without pressing any button. Failure of the controller. Continuous sound when simultaneously pressing the up and down column buttons. Motor current protection resetting. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 9 2D Bar Code (GS1 DataMatrix) DI+PI=UDI Electrical Specification 1D Bar code GS1-128 (Serial number) Symbols Please address any potential queries to the authorized representative or directly to the Manufacturer BORCAD Medical a.s. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 10: Un-Packing Manual

    Do not unpack and/or accept the consignment if the package has any visual damage! If so, issue a complaint. Personal qualification The persons authorised for installation and putting into operation of the product, i.e. those trained by BORCAD Medical a.s. Personnel required for unpacking and installation Minimum two persons. Anticipated time Anticipated time of unpacking is 30 minutes, anticipated time of installation is 30 minutes.
  • Page 11 Cut the plastic bag open carefully; ideally around the bottom edge of the pallet and remove the polythene from the chair including the bottom of the box. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 12 It is recommended to put the back of the bed in an upright position (60°) in order to access the box easily. Take out all the equipment from the box item-by-tem. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 13 Remove the other container from the pallet, so that the upper part is free. Unscrew the medical chair (if it is included in the package). It is recommended to put the bed to a higher position. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 14 Remove the bedhead (if it is contained in the package) from the holder and then unscrew it. Cut off and pull out the tape, release foot section Remove plastic plug from the locking element. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 15 Remove the rest of the packaging material off the bed. Make sure that all wheels are facing forward alongside the longer part of the pallet and lock the wheels to move in a straight line. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 16 Put two polystyrene ramps to the edge of the pallet. Carefully push the bed so that it gets from the anchoring wedges to the pallet and then move it down using the ramps. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 17 Back-up battery connecting With regard to the possibility of unwanted switching on of AVE2 upon packaging, transport and unpacking, the battery is normally disconnected from the equipment. The following pictures show how to connect the battery. Remove the undercarriage cover.
  • Page 18 Insert the leg rest in the joint. Hold the controller and insert the lock pin with the other hand. Release the leg rest controller and secure the pin with a protective cap. Screw the cap with two Allen wrenches with blue Loctite. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 19: Overview Of Basic Parts

    CPR functions leg rest tray detachable bed-head (accessories) BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 20: Overview Of Basic Versions

    The medical product is categorised as energised electric installation, hence, fire can be extinguished by the following types of extinguishers. Do not use Powder extinguisher Foam extinguisher CO2 extinguisher Water extinguisher Halotron extinguisher BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 21: Function

    When using the delivery bed as part of the medical system, certain elements (medical devices) may interact. Before this application it is necessary to follow recommendations provided in the user manual of the delivery bed AVE2. 3.2 General instructions The birthing bed may only be operated where the conditions of valid standards for el.
  • Page 22: Intended Use

    The birthing bed is not intended for use as a general hospital bed. 4.3 Operator • Caregiver • Patient (based on individual patient status assessment by caregiver the patient can utilize dedicated device functions) BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 23: General Safety

    • Never use other than original accessories. All original accessories are marked with an original label containing the name and brand of the manufacturer, designation of the BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 24: Electromagnetic Compatibility

    Mobile RF communication device (including end use devices like antenna cables and external antenna) should not be used closer than 30 cm (12 inches) from any part of this bed AVE2, including cables specified by manufacturer. Otherwise this could lead to deterioration of functionality of this bed.
  • Page 25 IEC 61000-4-11 a 315° 0 % U ; 1 cycle and 70 % U ; 25/30 cycle Single phase: at 0° 0 % U ; 250/300 cycle BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 26 NOTE There are applied no deviations to requirements of IEC 60601-1-2 ed. 4 NOTE There are no known other measures for keeping the basic safety based on EMC phenomena. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 27: Mechanical Safety

    When setting the bed all four castors have to be in contact with the floor and the brakes engaged. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 28: Setting The Product

    Trained personnel must be also educated on appropriate height of infusion stand setting. • The training is provided by BORCAD Medical a.s. (manufacturer) representative or its authorized dealership. We recomend to set up the bed in to Welcome position: •...
  • Page 29: Transport, Braking

    During transport, pay attention to the leg rests, side rails, foot section and hand grips to prevent hitting doors and other objects. Otherwise you can cause injury to persons or damage to the device. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 30 • Be sure the castors are unlocked before transport (see section related to adjustment of the castors and transport of the bed). Move the bed only on the floors with suitable surface. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 31: Transport

    Before transport of the patient, always put the infusion stand to the lowest position, thus avoiding injury to the patient or damage of the equipment. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 32 All castors free, spinning around their axle The left front castor is locked in straight direction, the other castors are unlocked, turning around their axes. Symbols meaning of arrest control pedals: NEUTRAL STOP BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 33: Operating The Bed

    • In case of fuse replacement pay attention to correct selection of a fuse type. Fuses need not be replaced by an authorized BORCAD Medical a.s. service; it can be done by a qualified electrician. •...
  • Page 34: Adjustment Of The Handle In Inactive Position

    Siderails are intended to be a reminder, not a patient restraining device. The siderails should always be in a full upright position and latched when a patient is unattended. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 35: Side Control Panel

    7.2 Side control panel Warning: Caution Side control panel is located on both sides of AVE2 birthing bed. The panel can be operated by trained personnel only. Holding the button initiates setting the bed to anti-shock position Call the nurse...
  • Page 36 Lowering / lifting of Indication of mains the seat section connection Bed height adjustment CPR lever Lowering / lifting of the back section Battery status indicator BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 37: Manual Setting Of The Cpr

    Keep holding the lever during this movement. To reset the back section to its original position, use motoric positioning (activates via hand control, side control panel or side rail panel). BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 38: Electric Setting Of Trendelenburg

    To reset the back section to its original position, use motoric positioning (activates via hand control, side control panel or side rail panel). BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 39: Setting Of The Foot Section

    Do not overload the footrest due to the unsecured positon. It is necessary to always secure the part in the end positions (parking, active)! Foot section positioning (sliding in/out, lifting/lowering, inclination) always without load. Folding/unfolding Tilting up/down Lifting/lowering BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 40 7.5.1 Folding/unfolding In lowered position, the foot section can be folded and locked in two positions: a) Folded (the part is hidden under the seat section) b) Unfolded BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 41: Lifting/Lowering

    All listed positions of the foot section do not limit motoric setting of the birthing bed. Before loading the foot section, its correct locking should be checked by short moves to the sides. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 42 Pull the control lever with the right hand and at the same time grab the front edge of the foot section and push downwards to the desired position. Releasing the control lever locks the selected position. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 43: Tilting Up/Down

    Then push the foot section downwards. You can lock the foot section in any position by releasing the control lever. Uninterrupted push downwards results in low horizontal position of the foot section. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 44: Adjusting The Leg Rests

    Gynecologic position and in the Horizontal position to maximize the bed surface. Basic overview Leg rest tray Folding/unfolding of the leg rest tray Turning the leg rest sideways Tilting of the leg rest Leg rest tray height adjustment BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 45: Gynaecological Position

    For positioning pull „Turning the leg rest sideways“ control lever as on the picture below and with the same hand pull the leg rest in the desired position. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 46: Foot Rest Position

    Then with same hand pull the tray to the desired position. Releasing the control lever locks the tray in the position. To reset the tray to its original position, repeat the procedure reversely. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 47 The leg rests can be adjusted in the Foot rest position. For positioning pull „Turning the leg rest sideways“ control lever as on the picture below and with same hand pull the leg rest in the desired position. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 48 Then pull the „Tilting of the leg rest“control lever as on the picture below and with same hand pull the leg rest in the desired position. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 49: Maximum Bed Surface Position

    Then pull the „Tilting of the leg rest“control lever as on the picture below and with the same hand position the leg rest in the horizontal position. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 50: Turning The Leg Rests Sideways

    Then pull the leg rest to the desired position. Releasing the control lever locks the position. To reset the tray to its original position, repeat the procedure reversely. +60° BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 51: Leg Rest Tray Ajdustment

    Then with the same hand adjust the leg rest tray in the desired position. Releasing the control lever locks the leg rest tray in the position. To reset the tray to its original position, repeat the procedure reversely. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 52: Leg Rest Tray Height Adjustment

    With the other hand adjust the leg rest tray in the desired height. Releasing the control lever locks the leg rest tray in the position. To reset the leg rest tray to its original height, repeat the procedure reversely. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 53: Leg Straps

    Loosen the strap (1), place the mother’s-to-be calf comfortably in the bowl of the foot rest (2) and fasten the strap (3). BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 54: Backup Battery

    Battery charging mode is indicated by a green indicator LED on side control panel. Battery status indicator Light on: battery disconnected or defective Flashing: charging the battery Light off: charging completed BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 55: Ppb-068 Grab Handles, Detachable (R/L)

    For inserting, simply push-in the handle - the safety member will fix the handle in its appropriate position automatically. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 56: Ppb-013 Grab Handles, Non-Detachable, Foldable (R/L)

    For inserting, simply pull-out the handle - the safety member will fix the handle in its appropriate position automatically. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 57: Holder Of Infusion Stand

    The bed is interconnected to the other device by means of a cable (not supplied with the bed) that is connected easily by its end to the coupling. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631...
  • Page 58: Exchangeability Of The Upholstery

    5 years of use. Date of the upholstery manufacturing can be found on the reverse side of the back and foot sections. 08-2016 BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 59: Accessories

    Lowering / lifting of the bed Lowering / lifting of the seat section Recommended position of the manual control unit: only on the left side of the seat section, see picture. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 60: Ppb-071 Supporting Bar

    Supporting bar is vertically locked up, it may be ejected from the holder in the upward direction! After having pulled out the cross-bar, it is necessary to drag both arresting elements to unlock them. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 61: Under Carriage Light

    Ensure that when the chair moves up and down, the surrounding objects do not collide or get caught with the Eurolath. The eurolath is used for placing the given accessories, such as armrests, hand control, infusion stand. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 62: Ppb-060 Rear Bed-Head

    “PRESS". 8.7 GKB-076 Doctor's chair (ergonomic) 8.8 ZK-05.X Doctor’s chair BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 63: Ppb-069 Front Bumper On Foot Section

    8.11 PPB – 070.01 Upper mattress (for 130 mm cushioning) The accessory is primarily used after post-partum; it increases the comfort of the resting mother. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 64: Ppa-010 Tray On The Swinging Holder

    The stainless steel tray is foldable on the swinging holder forth and backwards. It is possible to remove the tray from the holder in the forth position. Volume of the tray is 4,5 litres. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 65: Nurse Call

    Microphone indicator light Nurse call button Subsequently, the labouring woman can speak into a microphone with a set of speakers located in the rear part of the handle. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 66: Ppb-077 Front Bumper Fitted On The Leg Rest (R/L)

    With lifted back section (sitting position), the head rest also helps smaller patients to achieve optimal position on the edge of seat section. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 67: Bowl Ppb-034

    This flat, stainless-stell tray is in most cases used for tools necessary in post-delivery sutures. The tray can be parked below the seat section and, where required, you can place this tray onto existing and deeper trays (4,5l and 8l). BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 68: Ppb-088 Protractor Of The Seat Part, Ppb-098 Protractor Of The Back Part

    Caution Warning The protractor is only for information It is used to inform about the inclination of the back or the seat part of the delivery bed. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 69: Ppb-097 Bracket For Oxygen Bottles

    Contrary procedure To remove the oxygen bottle holder, the procedure must be carried out in the reverse order to placement on the delivery bed. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 70: S4-04642A Patient Card

    Use to hang the patient card in A4 format. Front bumper on the legs rest Patient card BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 71: Baby Plate

    When using the baby plate, the revolving bowl and the leg rest cannot be used together. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631...
  • Page 72: Ppb-074 Version With Mattress

    Then it can be used to expel the placenta and control it. If the baby plate is not used it must be removed from the bed and stored in the room. PPB-074 Version with mattress BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 73 (1) until you hear a click (2)! Then make sure the lock lever is in the down position (2) BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 74 (1). Then slide the baby plate out of the handles on the seat part of the bed towards you. (2). BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631...
  • Page 75: Care

    9 Care All BORCAD Medical a.s. products are designed for many years of service under the conditions of correct use. Please use your bed and its accessories carefully. Please follow these instructions for use and recommended operation and maintenance instructions. Avoid negligent handling. Do not, in case of a function failure, undertake any repairs and never use force! 9.1 Cleaning and disinfection...
  • Page 76: Cleaning The Bed

    8. Lift the chassis to the highest position, pull out the mains plug - remove the protective parts (right and left) and put them back in the original position after cleaning; 9. Lift the left chassis plastic cover and clean dirty areas. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 77 Make sure not to interfere with the electronics. After cleaning, first fit the right cover, then the left cover. 11. Infusion stand and holder; 12. Collecting tray (can be cleaned thermally, as stainless steel); 13. Casters and brake pedals. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 78: Recommended Disinfection (Disinfectants For Wiping)

    Surfanios premium Oxygen 0,5% Anios Oxy Floor Oxygen Incidin Active Oxygen Perform Warning! Do not use disinfectants with active substance: alcohol over 30%, active chlorine, iodine, aldehydes. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 79: How To Properly Coat Bed With Hospital Bedding

    We recommend covering exposed surfaces (seat section, foot rest) additionally with absorption gauze. Other examples of correct coating of the bed using hospital coatings before labour are shown below. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 80: How To Properly Coat Bed With Disposable Bed Sheets

    9.1.4 How to properly coat bed with Disposable bed sheets Special-shape disposable coating materials, designed specifically for AVE2 birthing bed, can also be used. Swing out the back section of the mattress and put the coating on it. Then return the back section to its place, lift the seat section and put the coating on it as shown below.
  • Page 81: How To Properly Coat Bed With Terrycloth

    9.1.5 How to properly coat bed with terrycloth You can also use special-shape terrycloth set directly onto AVE2 birthing bed. Swing out the seat section of the mattress and put the coating on it. Return the seat section to its place and lift the back section. Put the coating on it as shown below.
  • Page 82: Maintenance

    (loadless). If the measured value is below 12V (24V in case of measuring two battery cells together), it is necessary to replace the battery immediately. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 83: List Of Error, Malfunction And Generated Messages

    The bed emits intermittent acoustic signal Repair of the faulty Controller failure Call the service and one of the controller controllers is not functioning BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 84: Technical Parameters

    1,2 kg (2,64 lb) Weight of mattress (foot section) 2,6 kg (5,73 lb) Weight of mattress (pair of leg rest) Noise level Less than 65 dB (A) BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 85: Electrical Parameters

    Device class Classification of applied parts 100 VAC 50/60 Hz 110-120 VAC 50/60 Hz 127 VAC 50/60 Hz All possibilities of voltage – input 220-240 VAC 50/60 Hz BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 86: Attached Parts Of Product

    ČSN EN 60601-1 ed. 2:2007, by which they meet the requirements on classification of Type B attached parts. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 87: Transport Instruction, Storage Instruction

    Operation: 10°C – 40°C, 30% – 75%, 795 hPa – 1060 hPa Storage and Transport: -10°C – 50°C, 20% – 90% (non-condensing), 795 hPa – 1060 hPa BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 88: Service And Repairs

    Fax: 558 668 087 e-mail: service@borcadmedical.com For repairs, please use only the original BORCAD Medical a.s. spare parts and the services of trained service technicians licensed to repair products of this type. Unprofessional repairs shall be considered as a substantial breach of the warranty...
  • Page 89: Environmental Protection And Liquidation Of The Product

    11 Environmental protection and liquidation of the product Company BORCAD Medical a.s. is aware of the important role that the protection of the environment plays for future generations. Therefore, it focuses closely on the development, innovation and production of such technologies and materials that preserve the environment.
  • Page 90: List Of Available Equipment

    12 List of available equipment N (Basic versions): PPB-AX Birthing bed AVE2: Standard leg rests + leg straps (pair) Backup battery Sliding foot section including the upholstery Seamless mattress Manual CPR Infusion stand holder on both sides of the bed...
  • Page 91 PPB-006.07 Upholstery of the footrest PPB-007.03-X PPB-007.03 Upholstery of the right leg holder PPB-008.03-X PPB-008.03 Upholstery of the left leg holder N (Nurse Call) PPB-046 Nurse Call PPB-047 NO Nurse Call BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...
  • Page 92 N (Footrest) PPB-076 Set of leg rests (US) PPB-075 Set of leg rests (outside US) N (Stickers) PPB-030.05 Sticker BORCAD Medical PPB-YXZ0 Nameplate BORCAD Medical PPB-030.80 Warning stickers N (Mandatory selection) PPB-067-X Supporting bar PPB-062 Hand controller PPB-064 Eurobar PPB-065 Without eurolath PPB-070.01...
  • Page 93 Sliding stainless catch basin 4.5l PPB-034 Stainless steel bowl - 8 l PPB-009 Holder N (Choosing castors) PPB-078 Standard castors (150mm) PPB-079 Castors with longer pin (150mm) BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 631 www.borcadmedical.com...