1) Unscrew 4 screws from the back side of the foot control (fig. A)
2) Lift the right plastic cover of the undercarriage (please see chapter Cleaning
and disinfection).
3) Remove the protective cap at the bed base (fig. B), then press the button
under the cap by a non-metallic, non-conductive object so that the surface is
not scratched.
4) Then push on the button placed on printed circuit of the foot control (fig. C).
Small diode placed on the printed circuit will flash once. The main big diode
of the control will flash three times.
5) After the diode flashes, press the button placed on the printed circuit of the
receiver at the bed base (fig. D) again and cover the button by the protective
cap.
6) Now the correct connection is set up.
7) Screw the cover of remote control back (fig. A)
8) Set up the chair in standard position and test function of wireless control.