BORCAD Medical AVE PPA-AX30 Instruction Manual page 56

Birthing bed
Table of Contents

Advertisement

The foot
controller or
some of its
buttons are not
working.
The manual
controller or
some of its
buttons are not
working.
The massage
element does
not work.
The massage
element does
not work.
segment is defective.
The bed does
not react to
commands of
the controllers
BORCAD Medical a.s.
739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087
The cable from the
controller to the bed
is not inserted deep
enough into the bed
connector.
The cable from the
controller to the bed
is not inserted deep
enough into the bed
connector.
One of the safety
fuses of the
massage segment
has burnt out.
The button to start
the massage
The control box
reports an error.
Make sure the cable
from the controller is
inserted into the bed
connector properly.
Make sure the cable
from the controller is
inserted into the bed
connector properly
Check the fuses in the
massage segment (see
the procedure in the
service manual)
Check, if the massage
can be triggered
manually by connecting
the wires in the
compressor (see the
procedure in the
service manual)
Reset the control box.
CONTROL BOX
RESET: Press and hold
buttons for the
movement of the seat
section up and down.
An intermittent acoustic
signal must be heard.
Hold the buttons until
the acoustic signalling
stops. Repeat the steps
if acoustic signal is not
produced.
Check, if the bed
works with a different
remote foot
controller. If so,
replace it.
Check, if the bed
works with a different
manual controller. If
so, replace it.
Replace the burnt
fuses.
It the massage can
be started in the
described test, the
compressor is
functional. You just
need to replace the
button which triggers
the massage.
Diagnose the control
box. Replace the
faulty component.
56
www.borcadmedical.com