BS 7176 in the Medium hazard category. The upholstery meets the requirements of BS EN 1021-1, BS EN 1021-2, BS 5852 – resistance to ignition sources 5 BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 4
IEC 529 Safe operational load Safe patient load Product weight Battery marking Producer Date of production BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 5
Use special collection points for old electronic appliances.“ The medical device includes an RF transmitter which emits non- ionizing electromagnetic radiation. Cardiopulmonary resuscitation BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 6
International Special Committee on Radio Interference Any potential queries address please to the authorized representative or directly to the producer BORCAD Medical a.s. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087...
1 Un-packing manual Carefully cut through prospect and manual pockets and remove them out. Carefully cut through cover following bottom edge of palette. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 8
Carefully cut through foil witch hold electrical cable and put into electricity. Cut through the tape around the box and remove the trapeze. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087...
Page 9
Raise the bed back into the position 60 °, the driver is attached to the drawbar engine back under the bed. Cut the tape on the tray and gradually remove all the accessories. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 10
Remove the box with a bowl if it is included. All wheels rotate in the direction to the longer end of the palette and lock wheels in the direction. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 11
Put polystyrene runs to the edge of the palette. Gently push down on the bed in the direction of the arrows. Go down from the palette. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087...
Page 12
It is necessary to put the plastic cover into original position after the connector is plugged in. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
13. security screw of holder of leg rest 14. protective ledge of the plastic cover of chassis 15. control light for power supply 16. holder for infusion stand BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 15
(medical devices) may interact. Before this application it is necessary to follow recommendations provided in the user manual of the delivery bed. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 16
Medical Instrumentation the product may only be operated by persons with adequate qualification. Performance of servicing works: BORCAD Medical a.s. authorized service may only responsibly evaluate the safety technological features if the operators, repairs, alternations, etc., had been performed only at the producer authorized locations and the product is operated in full conformity with the instruction manual.
Page 17
Interference is possible in the vicinity of devices marked with the following label. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 18
IEC 61000-3-2 supplies buildings used for residential Voltage fluctuations/ purposes. flicker emissions applicable IEC 61000-3-3 BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 19
2 minutes of operation followed by 18 minutes of stand by. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087...
Training will be performed by a representative of BORCAD Medical a.s. (manufacturer) or its official dealer Maximum weight of the patient 210 kg.
Page 21
If you fail to do so, you can cause injury to persons or damage the device. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087...
In case of fuse replacement pay attention to correct selection of fuse type. Fuses need not be replaced by an authorized BORCAD Medical a.s. it can be done by a qualified electrician.
If the remote control is coupled with the bed for the second time and it was already coupled with this bed before, then the remote control is coupled for a new position (unless the reset occurred). BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 26
8) Set up the chair in standard position and test function of wireless control. Fig. A Fig. B Fig. C Fig. D BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 27
Remote control does not work if the birthing bed is not connected to the power network. It is recommended to replace the batteries every two years! BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 28
Back part of the wireless control. Use screwdriver to enable the four corner screws holding the back cover. Remove the cover, giving you access to batteries. Change batteries. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 29
Back up time - it is assumed 1-2 weeks depending especially on local temperature. The warmer temperature is the longer backup time you have. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Keep holding the lever during this movement. When the bed reaches horizontal or Trendelenburg position, release the lever. The bed locks safely in this position. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
If you wish to fluently pull out or push in the foot section completely, just keep the control pressed. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 32
When unfolding the foot section other person(s) shall not interfere. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
After releasing the upper securing lever the upper tray of the rests may be secured as follows: Footrest vertical positioning angle -90°/+45° Footrest horizontal positioning angle -16°/+48° Locked Loose BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 34
Leg rests may be set to so called “nonfunctioning” position along the leg part BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087...
Page 35
Motoric movement of the seat part requires proper visual control of the position of underfoot supports. Be extremely careful when adjusting the particular positions and, most importantly, avoid the positions as follows: BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
In the interest of easy removal/installation, ensure perpendicular position of the top part in relation to the bottom part of the leg support as shown in the following figure. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 37
It is recommended to match the colour marks on the fixed and removable parts so that the foot rests are not interchanged. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
When driving down with the bed must be leg holders in upper position. When are leg holders tilt down can come to the collision with the bed frame or with the floor! BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
After adjusting the leg support, tighten the small screw back to its position. Fig. 1 Fig. 2 Fig. 3 Fig. 4 Fig. 5 Fig. 6 BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
To clean grab handle girdles, you can use a weak solution of soap, alcohol cleaner or isopropyl alcohol. Do not use abrasive cleaners (such as liquid sand) and organic solvents not listed above. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
2. By repeated pressing of the spot higher massaging speed is activated. 3. Third pressing stops the massaging function. If massage is not turned off, it is stopped automatically after 15 minutes. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
The bed is interconnected to the other device by a cable (not supplied with the bed), end connector of which is gently connected to the coupling BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087...
In case of ordering special design of the existing mattress, the back and seat sections can be equipped with openable zippers to allow better access to the PUR foam (PPA-370). BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
3.1 PPA-29-X Supporting upholstered bar Maximum load of the trapeze bar is 70 Kg. 3.2 PPA-33 Large stainless steel bowl - 10 l BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
It is also necessary to take extra care when manipulating with the bed to avoid a conflict between the infusion stand and surrounding objects. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
3.13 PPA-00.37 Side support without upholstery (pair) 3.14 PPA-00.35-X Side support with upholstery (pair) 3.15 PPA-00.36-X Double side support with upholstery (pair) BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 51
Before using, the side-guard function must be checked to ensure firm attachment. During regular service inspection, the side guards must be checked and duly adjusted. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
4 Care All BORCAD Medical a.s. products are designed for many years of service under the condition of correct use. Please use your bed and its accessories with care. Please follow these instructions and the herein described operation procedures. Avoid negligent handling. Do not, in case of function failure, undertake any repairs and never use force! 4.1 Cleaning and disinfection...
Page 53
It is always necessary to follow the manufacturer’s instructions on individual cleaning and disinfecting agents. BORCAD Medical a.s. is not responsible for damage to the cushioning (e.g. cracking, paint flaking etc.), which would result from improper use of disinfectants. The warranty terminates in such cases.
Page 54
The bed must not be used without a properly fitted plastic undercarriage cover! Base or chassis covered in high strength plastic or metal with epoxy paint. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Pair the transmitter with paired with the remote foot working. The the receiver. receiver. controller. If so, battery icon is replace it. flashing orange. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Page 56
Hold the buttons until the acoustic signalling stops. Repeat the steps if acoustic signal is not produced. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
Nominal value of the fuse in the control unit (230 V) T 1,25 AL_250V Nominal value of the fuse in the control unit (100/120 V) 2,5 AT BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
(0/+25º) Leg part height adjustment (0/225 mm) Aerial turning of leg part: (0,60,90) Spatial adjustment of leg rest using two ball joints. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
ČSN EN 60601-1 ed. 2:2007, by which they meet the requirements on classification of Type B applied parts. BORCAD Medical a.s. 739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087 www.borcadmedical.com...
The warranty period for the product is 3 years provided the regular annual service checks are made. For all medical equipment produced by the BORCAD Medical a.s. company, the operator must ensure regular technical safety inspections according to the producer’s recommendation. The inspections must take place at least 1x per year or after the carrying out of any repair of the medical equipment or after any malfunction of the electrical system of the medical equipment.
Page 62
For repairs only use original BORCAD Medical a.s. spare parts and the services of trained service technicians licensed to repair products of this type. Unprofessional repairs shall be considered material breach of warranty...
5.8 Environment Protection Estimated service life of this medical device is 15 years. BORCAD Medical a.s. is aware of the important role that the protection of our environment plays for future generations. The materials of this product are environmentally compatible. It does not contain hazardous substances on the basis of cadmium, mercury, asbestos, PCB or CFC.