Electrical safety and EMC:
 Since the unit is powered from the network, it may cause interference
with sensitive equipment due to generation of electromagnetic field. In
order to reduce the influence of undesired electromagnetic effects, the
bed was designed in compliance with ČSN EN 60601-1-2. To prevent
the occurrence of these problems the bed must be used in compliance
with this manual.
 If there is a fluctuation in the voltage range (+/- 20 V) then it is
necessary to connect the product to voltage regulator. Otherwise the
electronics can be damaged.
 To avoid risk of electric shock, this chair must be connected to power
main with Protective earthing
 The chair must be positioned in such way, so if its necessary could be
immediately disconnected from the power main
 Modifications to this chair is forbidden
 Before plugging check the power cable for mechanical damage.
 Before plugging in check the voltage of the network for conformity with
information on the type label of the bed. A plugged-in bed is indicated by a
green control light on the foot of the bed.
 Power cable of the bed must not be connected through an extension cord
but directly into the power socket.
 Avoid hanging the power cable over moving parts of the bed. When
accumulators are in use, the cable may be damaged!
 No input cable of this or other devices can lead over the bed and /or over
the patient.
 Within the space of 1.5 m, within the patient space, only medical equipment
meeting standard EN 60601-1 or medical equipment with certification
according to IEC may be used.
 Birthing bed is not intended for use in areas enriched with oxygen
(concentration higher than 25% or partial pressure reaching 27.5 kPa).
 The birthing bed is intended only for use by medical professionals.
 The birthing bed may cause radio interference, which can have effects
on the operation of nearby equipment. It may be necessary to take
measures to reduce these effects, such as redirection, relocation or
shielding of the device.
 Interference is possible in the vicinity of devices marked with the
following label.
BORCAD Medical a.s.
739 45 Fryčovice 673, tel.: +420 558 640 611, fax: +420 558 668 087
17
www.borcadmedical.com