Raman Effect - Agilent Technologies 1260 Infinity User Manual

Fluorescence detector
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

1
Introduction to the Fluorescence Detector

Raman Effect

Raman Effect
The Raman effect arises when the incident light excites molecules in the
sample which subsequently scatter the light. While most of this scattered light
is at the same wavelength as the incident light, some is scattered at a different
wavelength. This inelastically scattered light is called Raman scatter. It results
from the molecule changing it's molecular motions.
Figure 5
The energy difference between the incident light (E
light (E
state (i.e. getting the molecule to vibrate, E
the Raman shift.
E
Several different Raman shifted signals will often be observed; each being
associated with different vibrational or rotational motions of molecules in the
sample. The particular molecule and its environment will determine what
Raman signals will be observed (if any).
A plot of Raman intensity versus Raman shift is a Raman spectrum.
14
Raman
) is equal to the energy involved in changing the molecule's vibrational
s
= E
- E
v
i
s
) and the Raman scattered
i
). This energy difference is called
v
Agilent 1260 FLD User Manual

Advertisement

Table of Contents
loading

Table of Contents