Sampling System Installation

Type 40 Probe Tube and Filter Unit Installation

Before the Installation
Figure 1
Type 40 Probe Tube
L1 = 500/1000/1500 mm
(Dimensions in mm)
Type 40 Probe Tube and
Filter Unit Installation
CI/ACX-EN Rev. D
CAUTION!
The weight of the probe tube with filter unit amounts to approx. 18–20 kg! Two
persons are needed for transportation and mounting!
Observe the "Piping Plan" in the drawings set.
Make sure that the wall tube is installed at the extraction point (see page 15).
Special Steel Material No. 1.4571 (max. 500 °C)
Step
Action
1
Screw the probe tube into the internal thread of the filter unit.
2
Insert the pre-assembled probe tube with filter unit in the wall tube and
screw the flange of the filter unit to the flange of the wall tube. Use the
green seal from the accessories pack to seal the space between the
flanges of wall tube and filter unit.
3
Mount the heating sleeve or the ring heater on the filter unit.
4
If applicable, install the compressed-air hoses between the filter unit and
the back-purging unit (see page 39).
ACX Analyzer System Commissioning Instruction
L1
16
G3/4
25