Probe 2 Installation

Probe 2 Delivery Form
Assembly of the
Ceramic Inlet Filter
Figure 8
Probe 2:
Ceramic Inlet Filter
CI/ACX-EN Rev. D
Probe 2 is supplied in various partially pre-assembled component parts:
• Gas sampling probe with flange and internal heating rod
• Ceramic inlet filter (inner filter)
• Installation set for mounting the ceramic inlet filter (4 bolts M12 x 70 with nuts,
spring washers and washers)
• Harting connector, degree of protection IP55
• Protective box (option), degree of protection IP54
CAUTION!
Danger of breakage! The ceramic inlet filter of probe 2 is fragile.
First of all, assemble the ceramic inlet filter as shown in Figure 8. Please note that the
compression spring 4 has to be compressed by approx. 15 mm.
7 6
5
1
Filter Element
2
Sealing Gasket
3
Bush
4
Compression Spring
5
Pressure Disk
6
Bush
7
Screw
ACX Analyzer System Commissioning Instruction
3
4
2
1
2
Continued on next page
33