ABB ACX Commissioning Instruction page 13

Advanced cga solutions, analyzer system for emission monitoring, cement applications and process measurement
Table of Contents

Advertisement

CI/ACX-EN Rev. D
The gas sensor installed in the analyzer cabinet is not factory calibrated; it is
inoperable without calibration. Calibration of the gas sensor is the responsibility of
the operator.
Installation, commissioning, parameterization, operation, signal evaluation and
maintenance of the supplied gas warning center are the responsibility of the
operator.
WARNING!
If the above-mentioned instructions are not observed or the hydrogen monitoring of
the analyzer cabinet is installed incorrectly, a hydrogen explosion may occur in the
event of a malfunction.
Operation of this safety device should be checked during commissioning and at
regular intervals (min. 1 time a year).
ACX Analyzer System Commissioning Instruction
13