CI/ACX-EN Rev. D
•
The gas sensor installed in the analyzer cabinet is not factory calibrated; it is
inoperable without calibration. Calibration of the gas sensor is the responsibility of
the operator.
•
Installation, commissioning, parameterization, operation, signal evaluation and
maintenance of the supplied gas warning center are the responsibility of the
operator.
WARNING!
If the above-mentioned instructions are not observed or the hydrogen monitoring of
the analyzer cabinet is installed incorrectly, a hydrogen explosion may occur in the
event of a malfunction.
Operation of this safety device should be checked during commissioning and at
regular intervals (min. 1 time a year).
ACX Analyzer System Commissioning Instruction
13