Emerson Rosemount Analytical X-STREAM Enhanced Series Short Form Manual page 35

Gas analyzers
Table of Contents

Advertisement

Instruction Manual
HASXEE-IM-HS
05/2017
Special Performance Specifications for Gas Purity Measurements (ULCO & ULCO
Detection limit (4 σ)
Linearity
1 2
Zero-point drift
Span (sensitivity) drift
Repeatability
Response time (t
Permissible gas flow
Influence of gas flow
Maximum gas pressure
Influence of pressure
– At constant temperature
– With pressure compensation
Permissible ambient temperature
Influence of temperature
(at constant pressure)
– On zero point
– On span (sensitivity)
Thermostat control
Related to full scale
1
Constant pressure and temperature
2
Within 24 h; daily zero calibration requested
3
Tab. 1-6: Special Performance Specifications for Gas Purity Measurements (Low Ranges)
Note 1!
Not all data listed are applicable to all analyzer versions (e.g. 60 °C [140 °F] thermostatically
controlled box is not available for electrochemical and trace oxygen, nor for
Note 2!
For NDIR/UV/VIS measurements, take into account that
• sample gas may diffuse or be released by leakages into the analyzer enclosure
• if existent in the analyzer surroundings, the component to be measured may enter the enclosure.
Concentrations then may increase inside the enclosure. High concentrations of the component
to be measured inside the enclosure may influence the measurement by unintended absorption,
which could cause drift of the measurement.
A remedy for this issue is to purge the housing with gas not containing the component of interest.
Emerson Process Management GmbH & Co. OHG
1.6 Measurement Specifications
1 2
1 2 3
1 2 4
1 2
)
7
90
1 2
10 11
5
8
+15 to +35 °C (59 to 95 °F) +5 to +40 °C (41 to 104 °F)
6
Within 24 h; daily span calibration recommended
4
Related to measuring value
5
Temperature variation: ≤ 10 K per hour
6
From gas analyzer inlet at gas flow of 1.0 l/min
7
0–10...< 50 ppm CO
0–5...< 50 ppm CO
< 2 %
< 1 %
< 2 % resp. < 0.2 ppm
< 2 % resp. < 0.2 ppm
< 2 % resp. < 0.2 ppm
< 10 s
0.2–1.5 l/min.
< 2%
≤ 1500 hPa abs. (≤ 7 psig)
≤ 0.1 % per hPa
≤ 0.01 % per hPa
< 2 % per 10 K resp. < 0.2 ppm per 10 K
< 2 % per 10 K resp. < 0.2 ppm per 10 K
none
Note! 1 psi = 68.95 hPa
Barometric pressure sensor is required
8
Whichever value is higher
9
Limited to atmospheric if internal sample pump
10
Special conditions apply to model XEFD
11
X-STREAM XE
)
2
2
9
9
9
9
9
60 °C (140 °F)
19 in instruments).
1
2
1-17