Tabletop & Rack Mount Analyzers - Emerson Rosemount Analytical X-STREAM Enhanced Series Short Form Manual

Gas analyzers
Table of Contents

Advertisement

Short Form Manual
HASXEE-SFM-HS
05/2017
A.2 Assignment of Terminals and Sockets
A.2.1 Tabletop & Rack Mount Analyzers
Signal
Pin
Channel 1, (+) 4 (0) - 20 mA
1
Channel 2, (+) 4 (0) - 20 mA
2
Channel 3, (+) 4 (0) - 20 mA
3
Channel 4, (+) 4 (0) - 20 mA
4
Channel 5, (+) 4 (0) - 20 mA
5
unused
6
unused
7
Output1 (Failure) NC
8
Output1 (Failure) NO
9
Output1 (Failure) COM
10
Output3 (Off spec) NC
11
Output3 (Off spec) NO
12
Output3 (Off spec) COM
13
Socket X1 - Analog outputs, relay outputs 1...5
(Assignment of screw terminals adaptor: see XSTA on next page)
Signal
Pin
Input 1
1
Input 2
2
Input 3
3
Input 4
4
GND for all digital inputs
5
unused
6
unused
7
Output 5 NC
8
Output 5 NO
9
Output 5 COM
10
Output 6, NC
11
Output 6, NO
12
Output 6, COM
13
Output 7, NC
14
Output 7, NO
15
Output 7, COM
16
Output 8, NC
17
Output 8, NO
18
Output 8, COM
19
Socket X4 - Digital I/O
(Assignment of screw terminals adaptor: see XSTD on next page)
Emerson Process Management GmbH & Co. OHG
A.2
Assignment of Terminals and Sockets
Pin
Signal
14
Channel 1, GND
15
Channel 2, GND
16
Channel 3, GND
17
Channel 4, GND
18
Channel 5, GND
19
unused
20
Output2 (Maintenance request) NC
21
Output2 (Maintenance request) NO
22
Output2 (Maintenance request) COM
23
Output4 (Function check) NC
24
Output4 (Function check) NO
25
Output4 (Function check) COM
Note!
Configuration of relay contacts as per standard
factory setting (NAMUR status signals). NAMUR
status signals are automatically configured Fail
Safe!
Pin
Signal
20
Input 5
21
Input 6
22
Input 7
23
Output 9, NC
24
Output 9, NO
25
Output 9, COM
26
Output 10, NC
27
Output 10, NO
28
Output 10, COM
29
Output 11, NC
30
Output 11, NO
Note!
31
Output 11, COM
The configuration illustrated
32
Output 12, NC
here is that of the first socket,
33
Output 12, NO
labelled X4.1.
34
Output 12, COM
Inputs 8-14 and outputs 14-
35
Output 13, NC
36
Output 13, NO
22, are on the second socket
37
Output 13, COM
(X4.2), if installed.
Pin
Pin
1
6
2
7
3
8
4
9
5
MOD 485/
MOD 485/
Pin no.
RS 232
2 wire
4 wire
1
Common Common Common
2
not used
not used
3
not used
not used
4
RXD1(+)
not used
5
D1(+)
TXD1(+) Common
6
not used
not used
7
not used
not used
8
RXD0(-)
not used
9
D0(-)
TXD0(-)
Connector X2 - IOIOI - Serial Modbus interface
(Assignment of screw terminals adaptor: see XSTA on next
page)
Signal
Pin
Pin
{
}
BR 2 fit a wire bridge here to apply analog
5
BR 2 signal in current mode to input 2
*)
4
unused
3
{
}
BR 1 fit a wire bridge here to apply analog
2
BR 1 signal in current mode to input 1
**)
1
alternatively set jumper P1 on electronics board XASI
*)
alternatively set jumper P2 on electronics board XASI
**)
Connector X5 - Analog inputs
(Assignment of screw terminals adaptor: see XSTI on next page)
X-STREAM XE
3
RXD
TXD
1
not used
2
not used
not used
not used
not used
DC 24 V Input (
19" analyzer)
1
2
Pin 1
Signal
9
Input 2 high (+)
Pin no.
8
Input 2 low (-)
7
Input 1 low (-)
1
6
Input 1 high (+)
2
3
6
Pin 8
other
Ethernet connector
for Modbus
1:
ME
2:
+ 24 V
3:
0 V (⊥)
Signal
TX+
TX-
RX+
RX-
not used
A-17