2.6.1 Installation - X-STREAM XEGK, X-STREAM XEGP
Signal inputs and outputs (basic configuration)
The number of signal outputs actually available varies according to the unit's configuration.
Analog signals
Relay outputs
Analog signals and relay outputs are located on a shared 25-pin submin socket X1,
or on an optional terminals adaptor XSTA (
Specification of analog signal outputs:
Specification of relay outputs 1–4:
Electrical specification:
Notes!
Consider the installation notes in section 2.7.
Technical data and installation information for further available interfaces are given in the
X-STREAM Enhanced instruction manual. Connector layouts are part of the appendix of this
manual.
Signal
Channel 1, (+) 4 (0)–20 mA
Channel 2, (+) 4 (0)–20 mA
Channel 3, (+) 4 (0)–20 mA
Channel 4, (+) 4 (0)–20 mA
Channel 5, (+) 4 (0)–20 mA
unused
unused
Output1 (Failure) NC
Output1 (Failure) NO
Output1 (Failure) COM
Output3 (Off spec) NC
Output3 (Off spec) NO
Output3 (Off spec) COM
Fig. 2-7:
Socket X1 - Analog Outputs & Digital Outputs 1-4
2-14
page 2-21; not for XEGK).
4(0)–20 mA; burden: R
Dry relay change-over contacts can be used as
NO or NC.
max. 30 VDC, 1 A, 30 W
Pin
Pin
1
14
2
15
3
16
4
17
5
18
6
19
7
20
8
21
9
22
10
23
11
24
12
25
13
Note!
Configuration of relay contacts
as per standard factory setting (NAMUR
status signals). NAMUR status signals
are automatically configured Fail Safe!
Short Form Manual
≤ 500 Ω
B
Signal
Channel 1, GND
Channel 2, GND
Channel 3, GND
Channel 4, GND
Channel 5, GND
unused
Output2 (Maintenance request) NC
Output2 (Maintenance request) NO
Output2 (Maintenance request) COM
Output4 (Function check) NC
Output4 (Function check) NO
Output4 (Function check) COM
Emerson Process Management GmbH & Co. OHG
HASXEE-SFM-HS