Concentration
Alarm Monitor:
LoLo Level:
Lo Level:
Hi Level:
HiHi Level:
Hysteresis:
States..
3.8.6

Backup the Settings

Save-Load
Local Backup..
Factory Defaults..
USB Backup..
USB Firmware Update..
Emerson Process Management GmbH & Co. OHG
3.8 Checking the Settings
On
50.000 ppm
100.000 ppm
400.000 ppm
600.000 ppm
10.000 ppm
You may turn the alarm function On or Off
separately for each channel ("Alarm Monitor").
It´s also possible to use some of the thresh-
olds only. In this case, set the not used to a
level outside the range limits (for this, "Lo"
and "LoLo" support entering negative values).
In case of an alarm, you may like to enter
the STATES submenu, to check which one
is triggered.
The most important settings have now been
checked and the unit is configured to suite
your needs.
A backup copy of these configuration data
can now be saved.
Press
until SETUP and then open SAVE-
left
LOAD.
This menu gives you the choice, to either
• make a local backup to a protected memory
area
• restore the factory default settings, or
• make a backup to an external USB device.