2.6 Electrical Connections

Only qualified personnel, observing all applicable technical and legal
requirements, may install these devices and connect power and signal
cables.
Failure to comply may render the guarantee void and cause exposure to
risk of damage, injury or death.
The devices may only be installed by personnel who are aware of the
possible risks. Working on units with screw-type terminals for electrical
connections may require exposure to energized components.
Wall-mountable X-STREAM analyzers are not fitted with power switches and
are operational as soon as they are connected to a power supply. A switch
or circuit breaker (IEC 60947-1/-3) must be installed on the premises. The
switch or breaker must be located near the analyzer, easily accessible and
labelled as a power supply cut-off for the analyzer.
Units with screw-type terminals must be de-energized by unplugging it or
operating the separate cut-off switch or circuit breaker, when working on
the power connections.
To avoid the risk of electrical shock, all units must be earthed. For this
reason, a power cable with protective earth conductor must be used.
Any break in the earth wire inside or outside the unit may cause exposure
to the risk of electrocution and is therefore prohibited.
Emerson Process Management GmbH & Co. OHG
2.6 Installation - Electrical Connections
ELECTRICAL SHOCK HAZARD BY INSTALLATION
X-STREAM XE
2-9