Figure 10 Calibration Example - Fluke Calibration 6020 Series User Manual

Calibration baths
Table of Contents

Advertisement

R0 = 100.000
ALPHA = 0.0038500
t
= 80.00°C
L
measured t = 79.843°C
t
= 120.00°C
H
measured t = 119.914°C
Compute errors,
err
= 79.843 - 80.00°C = -0.157°C
L
err
= 119.914 - 120.00°C = -0.086°C
H
Compute R0,
x
(
0 086
.
)
80 0
' =
R0
120 0 80 0
Compute ALPHA,
+
(
1 0 00385 120 0
.
ALPHA' =

Figure 10 Calibration Example

− −
x
.
(
0157
.
)
120 0
.
0 00385
.
.
− +
x
. )(
0157
.
) (
120 0 80 0
.
+
=
.
1 100 000 100115
.
x
1 0 00385 80 0
.
. )(
0 086
.
11 Calibration Procedure
Calibration Example
.
.
)
+
=
1 0 00385
.
0 0038387
.
63