4 Quick Start
Setting the Temperature
Press "EXIT" and the bath temperature will be displayed again.
E
Return to the temperature display
Bath temperature display
24.73 C
The bath will heat or cool until it reaches the new set-point temperature. Set the
heater switch to position "HIGH" to allow the bath to more quickly reach a
higher temperature. The "HIGH" setting may be necessary to reach higher tem-
peratures and control at high temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cutout should be correctly set to
prevent this from happening. See Section 9.8.
To obtain optimum control stability adjust the proportional band as discussed in
Section 9.7.