Table 2 Table Of Various Fluids And Their Properties - Fluke Calibration 6020 Series User Manual

Calibration baths
Table of Contents

Advertisement

Table 2 Table of Various Fluids and Their Properties

Fluid
Lower
(# = Hart Part
Temperature
No.)
Limit*
Halocarbon 0.8
–90°C (v)**
#5019
Methanol
–96°C (fr)
Water
0°C (fr)
Ethylene
–35°C (fr)
Glycol—50%
#5020
Mineral Oil
40°C (v)
#5011
Dow Corning
–40°C (v)**
200.5
Silicone Oil
Dow Corning
–35°C (v)**
200.10
#5012
Dow Corning
7°C (v)
200.20
#5013
Dow Corning
25°C (v)
200.50
Silicone Oil #5014
Dow Corning 550
70°C (v)
#5016
Dow Corning 710
80°C (v)
#5017
Dow Corning
66°C (v)
210-H
Silicone Oil
Heat Transfer Salt
145°C (fr)
#5001
*Limiting Factors — b - boiling point e - high evaporation fl - flash point fr - freeze point v - viscosity — Flash point test oc = open cup cc = closed cup
**Very low water solubility, ice will form as a slush from condensation below freezing.
Upper
Temperature
Flash
Viscosity
Limit*
Point
(centistokes)
70°C (e)
NONE
5.7 @ –50°C
0.8 @ 40°C
0.5 @ 70°C
60°C (b)
54°C
1.3 @ –35°C
0.66 @ 0°C
0.45 @ 20°C
95°C (b)
NONE
1 @ 25°C
0.4 @ 75°C
90°C (b)
NONE
7 @ 0°C
2 @ 50°C
0.7 @ 100°C
190°C (fl)
190°C
15 @ 75°C
5 @ 125°C
133°C (fl, cc)
133°C
5 @ 25°C
165°C (fl, cc)
165°C
10 @ 25°C
3 @ 135°C
230°C (fl, cc)
230°C
20 @ 25°C
280°C (fl, cc)
280°C
20 @ 25°C
232°C (fl, cc)
232°C
50 @ 70°C
300°C (fl, oc)
10 @ 104°C
302°C (fl, oc)
302°C
50 @ 80°C
7 @ 204°C
315°C (fl, oc)
315°C
50 @ 66°C
14 @ 204°C
530°C
NONE
34 @ 150°C
6.5 @ 300°C
2.4 @ 500°C
Freezing Point: The freezing point of a fluid is an obvious limitation to stir-
ring. As the freezing point is approached high viscosity may also limit
performance.
Pour Point: This represents a handling limit for the fluid.
Viscosity: Points shown are at 50 and 10 centistokes viscosity. When viscosity
is greater than 50 centistokes stirring is very poor and the fluid is unsatisfactory
for bath applications. Optimum stirring generally occurs at 10 centistokes and
below.
Specific Heat
Specific
(cal/g/°C)
Gravity
1.71 @ 40°C
0.2
0.810 @ 0°C
0.6
0.792 @ 20°C
1.00
1.00
1.05
0.8 @ 0°C
0.87 @ 25°C
0.48 @ 25°C
0.84 @ 75°C
0.53 @ 75°C
0.81 @ 125°C
0.57 @ 125°C
0.92 @ 25°C
0.4
0.934 @ 25°C
0.43 @ 40°C
0.45 @ 100°C
0.482 @ 200°C
0.949 @ 25°C
0.370 @ 40°C
0.393 @ 100°C
0.420 @ 200°C
0.96 @ 25°C
0.4
1.07 @ 25°C
0.358 @ 40°C
0.386 @ 100°C
0.433 @ 200°C
1.11 @ 25°C
0.363 @ 40°C
0.454 @ 100°C
0.505 @ 200°C
0.96 @ 25°C
0.34 @ 100°C
2.0 @ 150°C
0.33
1.9 @ 300°C
1.7 @ 500°C
8 General Operation
Thermal
Thermal
Conductivity
Expansion
(cal/s/cm/°C)
(cm/cm/°C)
0.0004
0.0011
0.0005 @ 20°C
0.0014 @ 25°C
0.0014
0.0002 @ 25°C
0.001
0.00025 @ 25°C
0.0007 @ 50°C
0.00028 @ 25°C
0.00105
0.00032 @ 25°C
0.00108
0.00034 @ 25°C
0.00107
0.00037 @ 25°C
0.00104
0.00035 @ 25°C
0.00075
0.00035 @ 25°C
0.00077
0.0003
0.00095
0.0014
0.00041
Bath Fluid
Resistivity
12
(10
-cm )
5 @ 25°C
1000 @ 25°C
10 @ 150°C
1000 @ 25°C
50 @ 150°C
1000 @ 25°C
50 @ 150°C
1000 @ 25°C
50 @ 150°C
100 @ 25°C
1 @ 150°C
100 @ 25°C
1 @ 150°C
100 @ 25°C
1 @ 150°C
1.7 Ω /cm
3
33