Weighing Range Data In Securedisplay [Weighingrangedatainsecuredisplay]; Automatic Zero Tracking In Dosing Cycle [Automaticzerotrackingincycle]; Maximum Tare Weight (% Of Max.) [Maxtareweight]; Initial Zero-Setting Device Limit Negative (% Of Max.) [Initialzerosettinglimitnegative] - Siemens SIMATIC ET 200SP Operating Instructions Manual

Technology module
Hide thumbs Also See for SIMATIC ET 200SP:
Table of Contents

Advertisement

6.5.16

Weighing range data in SecureDisplay [weighingRangeDataInSecureDisplay]

In legal trade operation, the software version of the HMI device must be recorded so that it can
be checked by an official verification officer.
6.5.17

Automatic zero tracking in dosing cycle [automaticZeroTrackingInCycle]

The parameter defines whether the zero tracking (if activated) is to be active only outside the
automatic dosing cycle (only in weighing step 0) or also inside the dosing cycle.
6.5.18

Maximum Tare weight (% of max.) [maxTareWeight]

The parameter is specified as a percentage of the "Maximum weight" parameter. All tare values
(semi-automatic, automatic, or preset tare) are checked for this limit and rejected if the limit is
exceeded.
6.5.19
Initial zero-setting device limit negative (% of max.)
[initialZeroSettingLimitNegative]
Requirement: The Initial zero-setting parameter is enabled.
Max. permissible negative deviation from the calibration zero point. The setting is made in % in
reference to the maximum weight of the weighing range 1.
With scales requiring official calibration, the sum of the negative and positive limit value may not
exceed 20%.
Example
Maximum weighing range = 100 kg
Maximum negative zeroing limit = 10%
→ Up to -10 kg (10% of 100 kg) can be set to zero by the function on power on.
See also
Initial zero-setting [initialZeroSetting] (Page 63)
6.5.20

Initial zero-setting device limit positive (% of max.) [initialZeroSettingLimitPositive]

Requirement: The Initial zero-setting parameter is enabled.
Max. permissible positive deviation from the calibration zero point if the "Initial zero-setting"
function is used. The setting is made in % in reference to the maximum load of the weighing
range 1.
With scales requiring official calibration, the sum of the negative and positive limit value may not
exceed 20%.
Technology module TM SIWAREX WP351 HF
Operating Instructions, 01/2020, A5E47521010-AA
Parameter assignment/addressing
6.5 DR03 Calibration parameters
65