Connecting prehensile cable
1) When using an Ottobock system hand, provide the system hand cable
with a threaded sleeve or clamp sleeve. When using an Ottobock cable
control hook, use 21A13=1, 21A13=2 or 21A14 Connection Pieces.
2) Shorten the cable of the body harness to the proper length. Screw the
cable to the threaded sleeve or clamp the clamp sleeve with the 736Y6
Clamping Tool. Connect the sleeve to the connection piece.
Completion
1) Rivet the eylet cable retainers to the proper position on the prosthetic
socket.
2) Alternative method: Provide eylet cable retainers with the 21Y203 Spacer
and screw them with 29C5=M4x9 Setting Nut and 503F3 Truss Head
Screw. Unlike rivet connection this connection can be loosened if need be.
4 Liability
Otto Bock Healthcare GmbH, hereafter referred to as manufacturer, as-
sumes liability only if the user complies with the processing, operating and
maintenance instructions as well as the service intervals. The manufacturer
explicitly states that this device may only be used in combination with com-
ponents that were authorised by the manufacturer (see instructions for use
and catalogues). The manufacturer does not assume liability for damage
caused by component combinations which it did not authorise.
The device may only be opened and repaired by authorised Ottobock
technicians.
5 CE Conformity
This device meets the requirements of the 93 / 42 / EEC guidelines for medi-
cal devices. This device has been classified as a class I device according
to the classification criteria outlined in appendix IX of the guidelines. The
declaration of conformity was therefore created by Ottobock with sole
responsibility according to appendix VII of the guidelines.
Ottobock | 17