Performing User Calibration; By Measurement (Single-Point Calibration) - Konica Minolta CA-410 Instruction Manual

Display color analyzer, probe + data processor
Hide thumbs Also See for CA-410:
Table of Contents

Advertisement

Performing User Calibration

1. By Measurement (Single-point calibration)

Operating Procedure
1
While the measurement screen is
displayed, press the [MENU] key.
The setting screen is displayed on the LCD screen.
2
Press the [UP] or [DOWN] key to select
[User Calib], and then press the [RIGHT]
or [ENTER] key.
3
Press the [UP] or [DOWN] key to select
[Measure and Register], and then press
the [RIGHT] or [ENTER] key.
4
The screen for selecting a probe is
displayed. Press the [UP] or [DOWN] key
to select the probe to register, and then
press the [RIGHT] or [ENTER] key.
5
The screen for selecting a calibration
channel is displayed. Press the [UP] or
[DOWN] key to select the channel to
register.
6
The screen for selecting a calibration
type is displayed. Press the [UP] or
[DOWN] key to select [Single-point
calibration], and then press the [ENTER]
key.
MENU
M e a s   C o n d
U s e r   C a l i b
T a r g e t
O p t i o n
S y s t e m
MENU
M e a s   C o n d
U s e r   C a l i b
T a r g e t
O p t i o n
S y s t e m
MENU
M e a s   C o n d
U s e r   C a l i b
T a r g e t
O p t i o n
S y s t e m
MENU
User Calib > Select CH
01
CH00
CH01
CH02
CH03
CH04
CH05
CH06
CH07
CH08
CH09
MENU
User Calib > Select Type
01
CH00
CH01
CH02
CH03
CH04
CH05
CH06
CH07
CH08
CH09
84
Trigger mode 0-cal AUTO
M e a s u r e   a n d   R e g i s t e r
C o p y   a n d   R e g i s t e r
E d i t   C o r r e c t i o n   C o e f
S e t I D
D e l e t e   C a l i b   D a t a
Trigger mode 0-cal AUTO
M e a s u r e   a n d   R e g i s t e r
C o p y   a n d   R e g i s t e r
E d i t   C o r r e c t i o n   C o e f
S e t I D
D e l e t e   C a l i b   D a t a
Trigger mode 0-cal AUTO
M e a s u r e   a n d   R e g i s t e r
01
C o p y   a n d   R e g i s t e r
02
E d i t   C o r r e c t i o n   C o e f
03
S e t I D
04
D e l e t e   C a l i b   D a t a
05
06
07
08
09
10
Single-point calibration