User Calibration - Konica Minolta CM-700d Instruction Manual

Hide thumbs Also See for CM-700d:
Table of Contents

Advertisement

User Calibration

You can perform calibration by using your own reference plate and calibration data instead of the white
calibration data. The calibration data for user calibration can be specified by connecting the instrument
to a PC and using the optional Color Data software "SpectraMagic™ NX".
You can select whether to use the user calibration data for measurement on the <User Cal.> screen of the
instrument.
[Setting Procedure]
1
Turn the instrument ON.
The <Calibration> screen is displayed.
• If a different screen is displayed, press the [CAL] button.
2
Press the [MENU] button.
The <User Cal.> screen is displayed.
3
Use the
button of the cross key to move the
cursor to "ON" and then press the [SAVE/
SEL] button.
Now you can perform white calibration using your own
calibration data.
N
ote
You need to prepare user calibration data and write it to the
instrument memory before you can set user calibration to
ON.
Calibration
E-31