Download Print this page

Verify The Installation - Keysight Technologies N9030B Installation Note

Signal analyzer, option bba analog baseband iq inputs upgrade kit
Hide thumbs Also See for N9030B:

Advertisement

Verify the Installation

1. Turn on the analyzer and wait for the instrument to fully boot up.
2. Verify the spectrum analyzer application loads and sweeps as expected. Press System, Show
Hard ware on the analyzer and verify that the Main board and the
themselves as:
— BBIQ
— BBIQ Interface
3. Press System, Show System on the analyzer to display a list of all installed options. You
should see N9030B-BBA.
Before performing adjustments and performance verification tests using the N7814A Calibration
and Adjustment Software for Keysight RF and Microwave Instruments, press MODE/MEAS, I/Q
Analyzer (Basic), OK. Press Input/Output, Select Input, I/Q, I/Q, I/Q Cable Calibrate... on the
analyzer and follow the on-screen directions.
If the newly installed hardware is functioning properly the cable cal should work fine. Press
MODE/MEAS, Spectrum Analyzer, Swept SA, OK. Press Input/Output, Select Input, RF.
Installation Note N9030-90077
Option BBA, Analog Baseband I/Q Inputs Upgrade Kit
BBIQ Interface Board
identify