Verify Hardware Installation
1. Verify the spectrum analyzer application loads and sweeps as expected.
2. Press System, Show, Hardware on the analyzer and verify that the Front End and Front End
Controller assemblies identify themselves as:
— Front End (Hw Id = 70, Part Number N9020-60314)
— Front End Controller (Hw Id = 75)
Verify Optional Functionality
1. Press Input/Output
2. Verify that there is a softkey labeled "External Mixer".
3. Press More, 1 of 2, Output Config.
4. Verify that there is a softkey labeled "Aux IF Out".
5. Press Aux IF Out and verify that there is a softkey labeled "Second IF".
Perform Preselector Characterization
(Only applies to analyzers with Option 508, 513, or 526)
1. Press System, Alignments, More, Advanced.
2. Press Characterize Preselector, Enter. The characterization will take several minutes.
3. Wait until the analyzer resumes sweeping.
Installation Note N9020-90230
Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328
37