Replacing A Fuse - Agilent Technologies XY Operator's Manual

Table of Contents

Advertisement

1 Disconnect the power cord from the power outlet and from the rear panel
2 Locate the fuse drawer on the rear panel. See Figure 21 on page 54 if
Figure 19
3 Insert a small screwdriver into the notch located at the left of the fuse drawer.
4 Twist the screwdriver to open and remove the fuse drawer. The fuse drawer
Figure 20
5 Remove the old fuse(s) and insert the new fuse(s).
6 Insert the fuse drawer into its receptacle in the XY autosampler.
Maintenance and Repair

Replacing a Fuse

Replacing a Fuse
A blown fuse (part number 2110-0303) may indicate the existence of another
problem in the instrument. If the replacement fuses blow, do not try others.
To change a fuse (part number 2110-0303), follow these steps.
receptacle.
necessary.
Locating the Fuse Drawer
contains one 2.0 A "T" Slo-Blo fuse (5 × 20 mm size) for a 100/120 V selection.
It contains two 2.0 A fuses for a 220/240 V selection.
Installing the Fuse
Post
Fuse
Fuse installation for 100/120 voltage
48
fuse drawer
Fuse
Fuse
Fuse installation for 220/240 voltage