1 Remove the cover plate from the front of the horizontal arm by removing its
2 Locate the hexagonal-shaped control rod and horizontal slider by looking
3 Using your finger, press on the control rod where it passes through the
4 Pull the needle foot, towards the front of the horizontal arm, as far as it will go.
Figure 5
horizontal slider
control rod
plastic plug
cover plate
needle foot
5 Position the vertical arm onto the horizontal slider. When viewed from the
Installation
Assembling the XY Autosampler
that the arm locking screw and safety shield have been removed and the
horizontal arm can move.
To install the vertical arm:
three screws.
down into the horizontal arm. See Figure 5.
horizontal slider. At the same time, pull the needle foot towards the front of
the horizontal arm. This causes the white plastic plug and control rod to move
forward slightly. When the white plastic plug is no longer flush with the front
of the horizontal arm, remove it and the control rod from the horizontal arm.
Installing the Vertical Arm—Part 1
front of the XY autosampler, the vertical arm's needle holder bracket is on the
right.