1 Screw the rinsing well (non flow-through rinse station) or flow-through
2 Attach the rinse station(s) to the support bar. If using only one rinse station,
Figure 9
optional rinse station
rinse station
3 Attach the support bar to the front of the XY autosampler using the knurled
Installation
Assembling the XY Autosampler
• flow-through rinse station — flow-through rinsing well.
Installing Rinse Station and Support Bar
rinsing well (flow-through rinse station) to the holder of the rinse station as
shown in Figure 9.
the default position for the rinse station is the left side of the support bar when
the bar is attached to the XY autosampler.
To attach tubing to the drain, locate the Tygon drain tubing. Remove the
locking collar from the drain fitting. Slide the tubing through the locking
collar. Attach the tubing to the drain fitting. Then screw the collar onto the
drain fitting. Place the other end of the tubing in a drain receptacle, located
lower than the tray.
If you ordered a flow-through rinsing well, remove the rinsing well and attach
the flow-through rinsing well in its place. Then attach tubing to the
flow-through timing well and connect the other end to a liquid source.
Mounting the Rinse Station
screws.
24
level-sensing rinsing well
drain
drain fitting
rinsing well
locking collar
tubing