1 Gently insert a small flat blade screwdriver into the SW1 selector, on the rear
2 Align the white dot with one of the indicated numbers. The unit ID is 10 plus
N O T E
Installation
Setting up the Communication Interface
Unit ID of the XY Autosampler
The unit ID identifies the XY autosampler to the UV-Vis ChemStation
software.
At the factory, the unit ID is set to 10. Since you will only connect a single XY
autosampler via the RS 232 interface at a time, there is no need for changing
this number.
However to change the unit ID do the following:
panel and turn it.
the selected number.
Unit ID set in the Agilent ChemStation Software
The default unit ID set in the Agilent ChemStation software is 10 and so
matches the default unit ID of the X-Y sampler.
To change the unit ID in the software:
1 Start your Agilent ChemStation and select Standard in the Mode drop
down list box.
2 Go to the Sampling System drop down list box and choose XY
autosampler as sampling system.
Your Agilent ChemStation software must be of revision A.06.0x or later in
order to have the capability to control the XY autosampler.
3 At the menu bar, go to Instrument and select Setup Sampling System in
the list that appears.
4 In the message box which pops up, change the number in the Unit-ID edit
box to the one appropriate for your setup.
Baud Rate
As a default, the baud rate selector is set to 0, indicating that the
XY autosampler is set for a baud rate of 19200.
The baud rate of the PC's RS 232 interface is set to 9600, and so you need to
change the setting for the SW2 selector to 1: