Agilent Technologies XY Operator's Manual page 23

Table of Contents

Advertisement

Figure 8
lock washer
needle guide
needle guide
1 Slide the needle into the tubing fitting.
2 Connect the transfer tubing from the dilutor, if installed, to the tubing fitting.
3 Slide the needle assembly into the needle holder on the vertical arm making
4 Secure the needle by tightening the needle retaining screw.
N O T E
Installation
Assembling the XY Autosampler
Needle Guide of Sampler Needle
Installing Needle Holder
To install the needle holder, slide it into place on the needle holder bracket
and secure with the supplied screw.
Installing Needle
To install the needle:
sure that the needle is centered over the needle guide.
Rinse Station and Support Bar Installation
The terms rinse station and wash station are used synonymously through
out this manual.
The support bar attaches to the front face of the XY autosampler. Its rear
contains mounting holes for rinse stations and its top contains mounting
holes for transfer ports and filler ports.
Rinsing Well
From left to right, the ports in a rinse station are the rinsing well,
level-sensing rinsing well, and drain. See Figure 9 for more details.
Before mounting the rinse station, you should make a decision as to whether
you want to use a
• non-flow-through rinse station — rinsing well,
or a
23