The table below provides a general description of all the system warnings that could
occur which would cause the system to pause. When a warning condition is
detected, a window will pop-up notifying the operator that the system is "paused"
because of a certain condition (below). When the condition has been corrected, the
"CONTINUE" button can be pressed to restart the system. If the problem cannot be
corrected, the "SHUTDOWN" button can be pressed.
Warning
Pump Flow Rate Under Range.
Pump #1-4 Not In Process
Page 64
LPX AUTOMATED SEED TREATER

PAUSED CONTROLLED WARNING

Possible Cause
Actual flow rate is under 80% of
target flow rate.
Valve of the liquid displayed
failed to divert to process when
requested
USC,
Check for empty supply tanks,
worn or obstructed hoses, etc.
that would cause a loss of liquid
flow.
Verify valve has diverted, if so
troubleshoot sensor, if not, check
air supply and signal to valve.