Once the three-phase adapter (current clamp) mode is set, it remains set even if
the ESC 2051 is disconnected from mains.
As long as the three-phase adapter (current clamp) mode is selected, „X"
symbol is displayed. There is no voltage present in Test socket 1 during
measurement.
Step 3
Switch on the main switch of equipment under test; it starts to run.
Step 4
Press the START key and keep it pressed, until displayed result is stabilized. Symbol
„ " indicates, that measurement is running.
If you need to lock the START key, proceed as follows:
Press ▲ key and keep it pressed. Press the START key.
Release both keys. Symbols „A" and „ " are displayed, indicating that locking of the
START key is activated and measurement is running.
Step 5
If locking of the START key is activated: measured value is automatically written to
the buffer after you cancel the START key locking by means of ▲ key.
If locking of the START key is not activated: last measured value is displayed and
automatically written to the buffer after you release the START key.
Switch off the main switch of equipment under test.

4.9. Tested equipment power and current

4.9.1. Power measurement of equipments equipped with mains plug

Step 1
Press the P
VA
Either result of last measurement or „---" will be displayed.
Step 2
Plug in the mains cable of equipment under test to Test socket 1.
Press the START key and keep it pressed. Symbol „ " indicates, that measurement is
running.
Copyright © 2004 ILLKO, s.r.o.
key and release it to select power measurement function.
Device power
---
43
ESC 2051