Example 2
You would like to send the measurement data from a Google email address to your email
address. The data collection interval of the MAG 8000 is 1 minute and the data collection
interval is 24 hours. You would like to receive the measurement data of all MAG 8000 devices
at the same time (06:00).
Step SMS to the Wireless Communication Module
1
Configuration;P3=+491722270333;
2
Configuration;P10=web.vodafone.de;
3
Configuration;P7=+4911111111;
P8=+4922222222;
4
Configuration;P52=8;
5
Configuration;P14=smtp.gmail.com;
P15=my.mag8000@gmail;P16=xxx;
P17=mag8000.dk@gmail.com;P42=465;
6
Configuration;P18=a@mail.com;
P19=b@mail.com;P20=c@mail.com;
7
Configuration;P55=1min;P56=24h;
8
Configuration;P47=2;
9
Configuration;P62=06:00;
Parameter P62 should be configured last to ensure precise data synchronization
1)
SITRANS FM MAG 8000 3G/UMTS add-on module
Operating Instructions, 03/2020, A5E03644134-AC
SMS response
123456H123 2017-09-12 13:25
Configuration: OK
123456H123 2017-09-12 13:26
Configuration: OK
123456H123 2017-09-12 13:27
Configuration: OK
123456H123 2017-09-12 13:28
Configuration: OK
123456H123 2017-09-12 13:29
Configuration: OK
123456H123 2017-09-12 13:30
Configuration: OK
123456H123 2017-09-12 13:31
Configuration: OK
123456H123 2017-09-12 13:32
Configuration: OK
1)
123456H123 2017-09-12 13:33
Configuration: OK
Commissioning
6.3 Configuration
Description
SMSC for SIM card (here Voda‐
fone Germany)
Don't continue if the response
SMS is not received
APN for SIM card (here Voda‐
fone Germany)
Configuring for which mobile
phone numbers you receive
alarm SMS
Selecting secured email for data
transmission
Configuring mailbox. The mail‐
box has to be created before‐
hand. For a Google mailbox pa‐
rameter P14 has to be SMTP
(not TLS/SSL).
Configuring mailboxes for recep‐
tion of measurement data
Configuring data collection inter‐
val and transmission interval
Configuring local time zone
(here GMT+2)
Configuring the synchronization
time of data transmission
45