17

COLD WEATHER PRECAUTIONS

Do not leave your analyser in a cold place overnight.
Electronic devices that become cold suffer when taken into a warm place.
Condensation may form affecting analyser performance.
Analyser sensors are affected by condensation or water. When this happens,
readings may display as "-" & sensors may be permanently damaged.
If you think your analyser is affected by condensation or water
ingress, leave running in a warm place with pump 'ON' sampling fresh air for
a few hours - Connect charger to avoid draining batteries.
If you still experience problems please contact KANE Customer Service.