3.6 Cleaning Solutions

Use the only cleaning solutions (or chemical equivalents) listed below for cleaning.
Generic Formulation
Sodium Hypochlorite
Glutaraldehyde
Hydrogen Peroxide
Cavicide®
CAUTION:-
Do not use the following cleaners, they will damage the parts being cleaned:
• Isopropyl Alcohol
• Quaternary Ammonium (such as Virex)
• Solvents (such as acetone)
CAUTION:-
Do not spray cleaning solutions directly on the Lullaby
CAUTION:-
Apply the cleaning solution with a clean sponge or cloth. Do not allow liquids to seep
into the heater housing.
CAUTION:-
Always dry the surfaces with a clean, damp, soft cloth to avoid scratches and remove
any residue of the cleaning solution left after cleaning.

3.7 Cleaning and Disinfecting Individual Components

Patient Probe (Reusable)
CAUTION:-
Avoid placing excessive strain on the probe lead. When cleaning, be careful not to
pull or bend the lead at the probe tip. Always remove the probe from the Lullaby
Warmer by grasping the plug at the control panel. Do not pull on the probe lead.
CAUTION:-
Do not apply cold sterilization or cleaning solutions to the probe connector.
Lullaby
Warmer
TM
2042376-001 Rev G ©2012 General Electric Company. All rights reserved. 27
Perform the system checkout procedure explained
in Chapter 2 before using the Lullaby
Maximum Concentration
0.5% Aqueous Solution
2%
6%
100% spray
TM
Warmer.
TM
Warmer.