Log Mode; Log Entries - ABB AWT420 Operating Instruction

Universal 4-wire, dual-input transmitter
Hide thumbs Also See for AWT420:
Table of Contents

Advertisement

30
AW T4 2 0 | U N I V E R S A L 4 - W I R E , D U A L- I N P U T T R A N S M I T T E R | O I/A W T4 2 0 - E N R E V. B
...9 Operation

Log mode

Log mode pages display logged information in the sequence
it occurred.
Log mode pages comprise:
• Calibration Logs: a history of calibration routines.
One log is provided for each sensor and is displayed only
if the sensor is fitted. Each log can store 15 entries that
are displayed in date order.
• Alarm Log: a history of alarm events.
• Audit Log: a history of analyzer activity.
• Diagnostics Log: a history of diagnostic events.
Event date
Audit Log
No .
Event
01
description
02
03
04
Alarm number/
icon* in
occurrence order
*Icons not displayed on Alarm Log or Calibration Log
Figure 27 Log page example (Audit Log shown)
Log icon
04-09-2019
10:03:25
Event
Date
Time
03:09:19
Power Failure
22:03:24
Power Recovery
23:06:19
14:17:03
Power Failure
15:05:19
02:21:54
Power Recovery
08:04:19
11:08:31
CAL
Event time

Log entries

Example Calibration Log entries with descriptions are shown in
Table 2. Example Audit Log entries together with a description
are shown in Table 3. The Diagnostics Log shows the history of
diagnostic messages that have been displayed in the
Diagnostic View – see page <?>.
Log entry
Description
Cal Failed
Calibration procedure failed due to low slope or sample
temperature error.
Cal Aborted
Calibration aborted manually by the user.
Cal Missed
Note. Sensor-type specific.
Table 2 Calibration Log entries
Log entry
Description
Power Failure
Power to the transmitter is lost.
Power Recovery
Transmitter restarted after a power loss.
In Config.
User in Advanced/Configuration mode.
Time/Date
User has changed date/time.
Changed
Daylight Saving
Time changed due to daylight saving.
Table 3 Audit Log entries

Hide quick links:

Advertisement

Table of Contents
loading