AW T4 2 0 | U N I V E R S A L 4 - W I R E , D U A L- I N P U T T R A N S M I T T E R | O I/A W T4 2 0 - E N R E V. B
...15
Calibration procedures
Turbidity Total Suspended Solids (TSS)
This section should be read in conjunction with Operating instruction OI/ATS430-EN.
Menu
Comment
Sensor Verification
See Turbidity TSS Sensor Verification – page 71 .
Turbidity Cal
See Turbidity calibration – page 72.
There are two possible calibration modes:
• 1-point
• 2-point
TSS Cal
See TSS calibration – page 76.
There are two possible calibration modes:
• 1-point
• 2-point
TSS Manual Cal
See TSS manual calibration – page 78.
Sample Collection
See In process calibrations – page 79.
Collection Complete
See In process calibrations page 79.
Restore Cal Defaults
Resets slope and offset values to default values.
Default