Calibration procedures
... pH/Redox/ORP calibration

1-Point automatic calibration

Note. Before starting the calibration ensure the pH buffers
are set to the correct values.
1 At the Calibrate level, press the
The Calibrate menu is displayed:
Calibrate
S1 : pH/Redox (ORP)
S2 : pH/Redox (ORP)
pH Buffers
Hold Outputs
Back
2 Use the
/
keys to select the sensor to be
calibrated, and press the
The pH calibration menu is displayed:
Calibrate
Sensor Cal
Temperature Cal
Edit Cal
Sample Collection
Collection Complete
Restore Cal Defaults
Back
key:
Select
key to confirm selection.
Select
3 Use the
/
keys to select sensor calibration and
press the
key to confirm selection.
The calibration type is displayed:
Sensor Cal
Calibration Type
1-Pt Auto
Next
4 Use the
key to edit the calibration type.
Use the
/
keys to select the required calibration
type and press the
key to confirm selection.
Press the
key to proceed to the next step.
Calibrate
Immerse in Buffer 1
ABB
4.0 pH
Abort
5 Place the sensor into buffer 1 and press the
key to perform the calibration. The calibration
process screen is displayed.
Calibrate
pH
4.00 pH
Tmp
25.0 °C
Settling-Please Wait
Abort
On completion the result screen is displayed.
• If the calibration passes, the slope and offset values are
displayed.
• If the calibration fails, the failure reason is displayed.
Note. The calibration can be canceled at any time during
the process by pressing the Abort
Edit
Continue
key.