Advertisement

AW T4 2 0 | U N I V E R S A L 4 - W I R E , D U A L- I N P U T T R A N S M I T T E R | O I/A W T4 2 0 - E N R E V. B

Calibrate

Menu
Calibrate
Exit
Select
Menu
Comment
S1: <Sensor Type>
Note. Displayed only if a sensor is fitted to slot 1.
S1: <Sensor Tag>
Access the sensor 1 specific calibration pages, refer to the Calibration section (page 55)
for specific routines.
S2: <Sensor Type>
Note. Displayed only if a sensor is fitted to slot 1.
S2: <Sensor Tag>
Access the sensor 1 specific calibration pages, refer to the Calibration section (page 55)
for specific routines.
pH Buffers
Note. Displayed only if at least one pH sensor is fitted and Measurement Type = pH.
Buffer 1
Set the type/value of buffer solution 1.
• Supported buffer solutions from table below/user-defined.
Used Buffer 1
Note. Displayed only if Buffer 1 Type = User Defined.
Set the user defined buffer characteristic curve using the five point linearizer table
(pH against °C).
Buffer 2
Set the type/value of buffer solution 2.
• Supported buffer solutions from table below/user-defined.
Used Buffer 2
Note. Displayed only if Buffer 2 Type = User Defined.
Set the user defined buffer characteristic curve using the five point linearizer table
(pH against °C).
Hold Outputs
Set to automatically hold current outputs and alarms whilst a calibration is being performed.
• Disabled/Enabled.

Buffer solutions

AWR126B026EN
ABB Capsule 4.01pH
AWR126B027EN
ABB Capsule 7.00pH
3KXA163000L0201
ABB Capsule 9.00pH
AWR126B028EN
ABB Capsule 10.00pH
AWR126B011EN
Technical 4.01pH
AWR126B013EN
Technical 7.00pH
AWR126B015EN
Technical 10.01pH
AWR126B017EN
DIN19266 1.679pH
AWR126B018EN
DIN19266 4.005pH
AWR126B019EN
DIN19266 6.865pH
AWR126B020EN
DIN19266 9.180pH
AWR126B021EN
DIN19266 10.012pH
AWR126B022EN
NIST 4.001pH
AWR126B023EN
NIST 6.881pH
AWR126B024EN
NIST 9.225pH
3KXA163000L0202
NIST 10.062pH
AWR126B025EN
Phth. Free 4.00pH
0400/110
ABB Sachet 4.01pH
0400/120
ABB Sachet 7.00pH
0400/130
ABB Sachet 9.18pH
User Defined 1
User Defined 2
Used to calibrate the sensor.
Note. Calibrate menus are sensor-specific – refer to the Calibration section (page 55) for
specific routines.
Access to the Calibrate menu is permitted via the Calibrate and Advanced levels or directly from
an Operator page using the Cal button.
37
Default
ABB Capsule 4.01pH
N/A
ABB Capsule 9.00pH
N/A
Disabled