CHAPTER A
GENERAL & TECHNICAL DATA
4

TECHNICAL SPECIFICATIONS

Generator
X-ray tube
Focal spot size
Total filtration
Anode voltage
Anode current
Cooling period
Exposure time
Duty cycle for height adjustment
SID
Magnification
CCD pixel size
Image pixel size
CCD active surface
Line voltage
Power input
Line harmonics
Max. permissible apparent
impedance of supply mains
Electrical classification
Fuses
Weight
Colour
Ambient temperature
Humidity
Original manufacturer
Planmeca Oy, Asentajankatu 6, FIN-00880, Helsinki, Finland
Phone: +358 20 7795 500, Fax: +358 20 7795 555, www.planmeca.com
A-4
Planmeca ProOne
TECHNICAL SPECIFICATIONS
Constant potential, resonance mode
high frequency 60 - 80 kHz
according to IEC 60601-2-7: 1998
Toshiba D-058SB
0.5 x 0.5 mm
according to IEC 60336:2005
min. 2 mm Al
60 - 70 kV ± 5%
2 - 7 mA ±10%
Automatically controlled
2 - 10 s as indicated ±10%
1:10 or 2 min continuous use followed by
18 min not in use
480 mm (19 in.)
1.22 - 1.29 (panoramic exposures)
33 μm
99/132 μm (selectable)
9 x 136 mm (0.35 x 5.35 in.)
100 - 240 V~ / 50 - 60 Hz
Continuous operation with intermittent
loading
800W intermittent: 1:20
(70 kV, 7 mA, 10 s exposure)
25W continuous
According to IEC 61000-3-2 class A
0.5 Ω (100VAC)
Class I, type B applied part
2 user replaceable fuses
F 8A H 250V
67 kg (148 lb)
White, RAL 9016
Operating +10°C to +40°C
storage ±0°C to +50°C
transport ±0°C to +50°C
15% - 85%
Technical Manual