Step 3: Adjusting the patient support table in x-direction

On the Configuration display, select the option Arm calibration. The Arm calibration
display will appear.
Touch the Side field on the Arm calibration display to move the C-arm to position 1. In
position 1, the C-arm is positioned parallel to the arm of the patient support table (see Fig.
24).
Figure 21
Use a locking pin to lock the C-arm in this position. To do this, insert the locking pin into
the hole in the C-arm top frame and make sure that it reaches the positioning hole in the
linear movement block.
C-arm top frame
Locking pin
Figure 22
C-16
Planmeca ProOne
ADJUSTING AND CALIBRATING X-RAY
1. Move C-arm to next position
2. Adjust C-arm position
BEAM
Technical Manual