10.0 Eliminating errors

Description
Device cannot be switched on
Battery does not recharge.
Charging symbol does not light
up despite proper connection to
the mains cable
„Vacuum low"
„Vacuum too high"
„Battery low"
System shut down
40
Possible causes
Battery completely empty
Fuse defect
Mains cable defect or not plugged
in properly
Mains cable or battery are
internally defect
Leakage
Blockage in the hose system
Filter/ overflow protection in hose
/ filter is blocked
Liquids sucked into pump
Excessively high vacuum applied
from the outside
Venting valve defect
Battery almost empty
Battery empty
Measures
Connect the battery recharger
to the device to recharge the
battery,
Note that the progress can be
checked in the upper left hand
corner of the display
Control fuses
Control mains cable
Device must be checked by the
manufacturer or an authorised
service partner
Check all the hoses and the canister
for leakages. Check whether the
pop-off valve is loose and fix it if
necessary..
Remove the blockage if necessary.
Remove the measuring channel
from the connection to the
canister. If the filter in the
measuring channel is blocked
exchange the hosing system
Check hose and filter.
If the filter is wet exchange
canister and filter
The device must be checked
by the manufacturer or an
authorised service partner
Check for correct hose
connections
Device must be repaired by the
manufacturer or an authorised
service partner
Connect the battery recharger
to recharge the battery. The
progress is displayed in the
upper left hand corner of the
display.
Connect the battery recharger
to recharge the battery. The
progress is displayed in the
upper left hand corner of the
display.