13.3 Guidelines and Manufacturer´s Declaration - Immunity
The ATMOS
C 051 Thorax is intended for use in the electromagnetic environment specified below. The custo-
®
mer or user of the ATMOS
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
46
C 051 Thorax should ensure that it is used in such an environment.
®
IEC 60601-
Compliance Level
Test Level
V
= 3 V
3 V
1
eff
150 kHz bis 80 MHz
E
= 3 V/m
3 V/m
1
80 MHz bis 2,5 GHz
Electromagnetic Environment -
Guidance
Portable and mobile communications
equipment should be separated from
the ATMOS
C 051 Thorax incl. the
®
cables by no less than the distances
calculated/listed below.
Recommended distances:
d = [ 3,5 / V
] √P
1
d = [ 3,5 / E
] √P
1
from 80 MHz to 800 MHz
d = [ 7,0 / E
] √P
1
from 800 MHz to 2500 MHz
Where „P" is the max. power in watts
(W) and D is the recommended sepa-
ration distance in meters (m).
Field strengths from fixed transmit-
ters, as determined by an electroma-
gnetic site (a) survey, should be less
than the compliance level (b).
Interference may occur in the vicinity
of equipment containing following
symbols.