Operating; Explanation Of The Display - Atmos Thorax C 051 Operating Instructions Manual

Table of Contents

Advertisement

4.0 Operating

4.1 Explanation of the display

Flow
Display of the actual
fistula
ATTENTION: This is the
Actual vacuum
current value. In the case of
Display of
a discontinuous fistula, the
value can decline temporarily
the actual
to zero although the fistula
vacuum value
still exists.
Start / Stop
therapy
Target vacuum
Display of the preadjusted
target vacuum to which the
pump adjusts
Keylock manual
Key lock activated
Flow displayed as bubbles
Each additional coloured bubble
represents an additional flow.
None:
0 - < 30 ml/min
Green:
30 - < 60 ml/min
Yellow:
60 - < 630 ml/min
Orange: 630 ml - < 1,00 l/min
Red: > 1,00 l/min to maximum
Up to 1,00 l/min the flow is
displayed in ml/min.
System information
Therapy progress
Changeover to the
graphic diagram
User settings
Setting button
Increases the
target vacuum
Setting button
Decreases the
target vacuum
Day/Night Mode
The ATMOS
C 051 Thorax has a day/night
®
mode, the device adjusts automatically to the
environmental lighting.
Under low ambient light conditions display has
dark background illumination.