Notes On Emc - Atmos Thorax C 051 Operating Instructions Manual

Table of Contents

Advertisement

13.0 Notes on EMC (Electromagnetic compatibility)
Medical electrical equipment is subject to special precautions with regard to EMC and must be installed acc.
to the following EMC notes.
Portable and mobile HF communication facilities can infl uence medical electrical equipment.
The use of other accessories, other converters and cables than those stated may lead to an increased
emission or a reduced interference immunity of the equipment or system.
13.1 Guidelines and Manufacturer's Declaration - Emissions
The ATMOS
C 051 Thorax is intended for use in the electromagnetic environment specifi ed below. The customer or
®
user of the ATMOS
®
Emissions Test
HF emissions acc. to CISPR 11
HF emissions acc. to CISPR 11
Emission of harmonics
acc. to IEC 61000-3-2
Emissions of voltage fl uctuations/
fl icker acc. to IEC 61000-3-3
44
C 051 Thorax should ensure that it is used in such an environment.
Compliance
Group 1
Class B
Class A
In compliance with
Electromagnetic Environment
- Guidance
The ATMOS
C 051 Thorax uses
®
RF energy only for its internal
function. Therefore, its RF emissi-
ons are very low and are not likely
to cause any interference
in nearby electronic equipment.
The ATMOS
C 051 Thorax is sui-
®
table for use in all establishments,
including domestic, and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.