8.0 Eliminating errors

The ATMOS C 341 Battery was subjected to a thorough quality control in the factory.
Nevertheless, if a problem may occur you can possibly solve it yourself.
Recharging and battery status
Error indication
Device cannot be
recharged.
LED on the power supply
and recharging unit is not
illuminated.
When recharging the
battery 100 % cannot be
achieved. The charging
time can take up to 14
hours.
Red LED of the battery
status indicator is lit up
although the battery is
fully charged.
Battery compartment
cover cannot be closed.
Possible cause
The plug from the
charging accessories is
poorly fitted.
Charging accessories
are defective.
Battery is not connected
properly.
Battery temperature too
high or too low.
Defective battery.
Defective fuse or
electronics.
Defective power supply
and recharging unit.
The mains plug is poorly
fitted.
Battery service life
is exhausted or the
battery is defective.
Wrong charging
accessories.
Defective battery.
Battery is not fitted
correctly.
Battery compartment
cover is installed
incorrectly.
Remedy
Check the connection to the
mains supply.
Exchange charging accessories.
Check the plug connections in the
battery compartment.
After prolonged use: Let the
device cool down.
Extreme ambient temperature: If
necessary, move the device to a
cooler or warmer location.
Exchange the battery.
Send in the device for repair.
Exchange the power supply and
recharging unit.
Check the connection to the
mains supply.
Exchange the battery.
Only use the provided charging
accessories or an original spare
part.
Exchange the battery.
Fit the battery correctly.
Mount the battery compartment
cover correctly according to
operating instructions.