Leakage Test - Atmos Thorax C 051 Operating Instructions Manual

Table of Contents

Advertisement

4.4 Leakage test

Following device startup the leakage test starts
automatically.
The hose attachment towards the drainage catheter
should already be sealed with a sterile plug when starting
up the device. Alternatively the thorax catheter can be
clamped near to the patient.
Do not clamp the ATMOS hose system.
If the leakage test is error-free, the message „Leakage
test OK" appears. Now you can remove the plug from
the hose. By pressing the
button you will reach
the main menu.
If the leakage test is faulty, the message „Leakage test
failed" appears. You now have the possibility to
a) repeat the test
b) abort the test
by pressing the respective buttons on the screen.
ATTENTION: If the leakage test is operated accordingly
the leakage must not be ignored. If the device has been
dropped it must not be operated. Send in the device
for repair. Treatment with defective equipment can
cause fatal injuries to the patient.
The intention of "abort the leakage test "is to skip the test
if an standard test under given conditions is not possible.