4.0 Operation
Fig. 12
Fig. 13
Fig. 14
Fig. 15
14
4.8
Information on the display
Actual vacuum
Target vacuum
Text line
Battery status
Alarm symbol
4.9
Display functions
4.9.1. Switch on
Switch on the ATMOS
the
I/O key.
The start screen is shown for 2 seconds.
Visualized flow display.:
Each additional bar represents an additional flow.
From 1 to 3:
From 4 to 6:
From 7 to 9:
4.9.2 Adjusting the vacuum
Adjust the target vacuum with the keys▼▲.
The basic setting is -20 mbar. The last selected value is stored
and is adopted as the target value during the next start up of
the device.
Depending on the selected vacuum unit, the target
vacuum can be set in 1mbar or 1kPa steps.
NOTE: 1kPa = 10mbar, therefore when kPa is selected, the
vacuum range is 1kPa to 10kPa.
4.9.3 Leakage test
Perform a leakage test (see chapter 3.3). If the
!
leakage test failed, please repeat the test.
Only exit this menu when you wish to start the
therapy.
(eg. pump on or off)
Display of airfl ow
Key lock
Service display Please call service
Alarm signal is suppressed
Actual flow
E 201 Thorax with
®
0,25 l/min (small)
0,25 l/min (medium)
3,80 l/min (large)