Information On The Display; Display Function; Adjusting The Vacuum; Leakage Test - Atmos S 201 Thorax Operating Instructions Manual

Hide thumbs Also See for S 201 Thorax:
Table of Contents

Advertisement

4.0
Operation
Fig 14
Fig 16
Fig 17
Fig 18
4.8

Information on the display

Actual vacuum
Target vacuum
Air flow
Display of airflow
Alarm signal
Service display
Alarm signal is suppressed
Pause (stop pumping)
Key lock
Battery status
Fig 15
4.9
Display functions
4.9.1. Switch on
Switch on the ATMOS S 201 Thorax with the
I/O key.
The start screen is shown for 2 seconds.

4.9.2 Adjusting the vacuum

Adjust the target vacuum with the keys▼▲.
The basic setting is -20 mbar. The last selected value is stored
and is adopted as the target value during the next start up of
the device.

4.9.3 Leakage test

Perform a leakage test (see chapter 3.3). If the
!
leakage test failed, please repeat the test.
Only exit this menu when you wish to start the
therapy.
Please call customer service
Flow displayed as bubbles:
Each additional dark colored bubble
represents an additional flow.
From 1 to 9:
0,25 l/min (blue)
from 10 to 12: 3,30 l/min (red)